UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT
Ref. No.
SOP/PD/206_01
STANDARD OPERATING PROCEDURE
FORCE TEN
®
FDX
Valid for:
2 years from
approval
Page 9 of 11
HE
Press SELECT key twice to display
force cell capacity of gauge
2
Does the displayed
force cell capacity
match with that of
force cell module
Apply test weight that is equal to
the displayed full-scale capacity weight
Press SELECT key
Used weight is lower than the full
capacity of the force cell module
[uuuuu] is displayed
Repeat auto-calibration with a
standard weight equal to the full
capacity of the force cell module
[nnnnn] is displayed
Used weight is higher than the full
capacity of the force cell module
Wait until device displays the applied
weight to indicate that calibration
process was successful
End
Repeatedly press SCROLL key until
correct force cell capacity is displayed
3
1
Yes
Yes
Yes
No
No
No